Features:
Core
Complete built-in module design ensures stable and reliable performance
Support storage of 168 hours trend table and graph review, 100 minutes waveform review, 500 groups NIBP review and 100
hours of parameter alarms events and associated parameter
Body
12.1 inch high-brightness TFT LED
Support display Max.8 waveforms
Printer
Built-in high-speed 50mm thermal printer (Brand: SEIKO, Japan)
Central System
Support wired connection to the central monitoring station.
Alarm
Three-level acousto-optic alarm
Sensor-off alarm
Paper out alarm
Support alarm review
Support alarm pause
OS
Multi-display mode to meet different clinical requirement, including Standard, Large Font, Trend Table, Full lead ECG, Oxy CRG
NIBP self-test mode: including overpressure test, static pressure test and air leakage test
18 types of Arrhythmia analysis and real-time S-T segment analysis and pacemaker detection
Drug calculation and titration table
Support multi-language display
Support online software upgrading by LAN
Interface
Transducer socket
ECG
SPO2
NIBP
IBP 1 / IBP 2
TEMP 1
TEMP 2
LAN / Central monitor system interface
The LAN port is for connecting the monitor to a central monitor system network
Input device interface
Knob (Standard)
Touch screen (optional)
Service Features
The Support Tool helps technical personnel to carry out configuration, upgrades and
troubleshooting on an individual monitor.back up the monitor settings.
The Service Mode is password-protected and ensures that only trained staff can access service tests and tasks. It includes Demo,
Language Setting, NIBP Calibration and Touch Screen Calibration.
Related Product
IBP accessories
Performance Specifications
Dimension and Weight
-Weight: 3.4kg(excluding accessories)
- Dimension: 375mm*330mm*180mm
Power Supply
-Voltage: AC100~240V, 50/60HZ, Power≤60W
Display
- 12.1" color TFT LED
- Resolution: 800*600 pixels
Battery
- Type: Rechargeable lithium battery: 12V/2200mAh
- Charge Cycle: ≥500 times
- Working time: 2 hours (optional larger battery for 4-5 hours)
Recorder (Option)
- Method: Thermal printer
- Paper width: 50 mm (1.97 in)
- Printing speed: 25/ 50 mm/s
- Trace: Max. 2 tracks
- Recording way: Real-time Recording, Review printing,
- Recording way: Real-time Recording, Review printing, Alarm recording
Alarm
- Level: Low, medium and high
- Indication: Auditory and visual
- Alarm volume adjustable
- Alarm pause time: 1min, 2min
- Parameter alarm type: Latch/ Unlatch
Input Device
- Touch screen (option)
- Knob (Standard)
- Keypad input (Standard)
System Output & Extensible
Interface
- Ethernet Network: 1 standard RJ45 socket
- With reference to RoHS Directive 2011/65/EU recasting
Trend & Reviewing
- Trend: 168 hours
- NIBP measurement reviewing: 500 groups
- Waveform review: 100 mins
- Alarm event: 100 hrs of parameter alarms events and associated parameter
Heart Rate
- Measure range:
Adult: 0bpm; 15 ~ 300 bpm Neo/Ped: 0bpm; 15 ~ 350 bpm
- Resolution: 1 bpm
- Accuracy: ± 1% or ± 1bpm, whichever is great
ECG
- Lead mode: 3/5 Leads, I, II, III or I, II, III, AVR, AVL, AVF, V
- Gain: 2.5mm/mV (×0.25), 5.0mm/mV (×0.5), 10mm/mV (×1), 20mm/mV (×2)
- Sweep speed: 12.5 mm/s, 25 mm/s, 50 mm/s
- ECG signal range: ±6 mV
- Accuracy: ±1bpm/ ±1%, whichever is greater
- Resolution: 1 bpm
- Leakage Current < 10 μA
- Baseline Recovery: < 5 s After defibrillation (MOR or Surgery mode)
- Bandwidth: Surgery 1 ~ 20 Hz Monitor 0.5 ~ 40 Hz Diagnostic 0.05 ~ 130 Hz
- Indication of Electrode Separation: Every electrode (exclusive of RL)
ST Measurement
- Range: -2.0 ~ +2.0 mV
- Range: -2.0 ~ +2.0 mV
- Other range: unspecified
- Resolution: 0.01mV
Respiration
- Method: Impedance between RA-LL, RA-LA
- Gain: ×0.25, ×0.50, ×1, ×2
- Respiration Rate: Adult: 0 ~ 120 BrPM; Neonatal / Pediatric: 0 ~ 150 BrPM
- Resolution: 1 BrPM
- Accuracy: ±2 bpm or ±2%, whichever is greater
- Apnea Alarm: 10 ~ 40 s
NIBP
- Method: Oscillometric
- Measure mode: Manual, Auto, STAT
- STAT mode cycle time: Keep 5 minutes, at 5 seconds interval
- Measure and Alarm Range:
Adult: SYS: 40 ~ 270 mmHg DIA: 10 ~ 215 mmHg
MEAN: 20 ~ 235 mmHg
Pediatric: SYS: 40 ~ 200 mmHg
DIA: 10 ~ 150 mmHg
MEAN: 20 ~ 155 mmHg
Neonate: SYS: 40 ~ 135 mmHg
DIA: 10 ~ 100 mmHg
MEAN: 20 ~ 110 mmHg
- Static pressure accuracy: ±3mmHg
- Resolution: 1mmHg
- Accuracy: Maximum Mean error ±5mmHg Maximum Standard deviation ≤8mmHg
Over pressure Protection: Dual protection via software & hardware
Temperature
- Technique: Thermistor probe (2.25K)
- Channel: Dual-channel, provide T1; T2; ΔT
- Measuring and Alarm Range: 0.0 ºC~ 50 ºC (32ºF~ 122ºF)
- Unit: Celsius (ºC), Fahrenheit ( ºF)
- Resolution: 0.1ºCor 1ºF
- Accuracy: ±0.1ºC(exclusive probe)
SpO2 (Digital Technic)
- Measurement Range: 0 ~ 100 %
- Resolution: 1 %
Accuracy: 70% ~ 100%: ±2 % 0% ~ 69%, unspecified;
Pulse Rate
- Measuring and Alarm Range: 20~300bpm
- Accuracy: ±3bpm
- Resolution: 1bpm
IBP (Option)
- Max Channel: 2
- Measurement way: Thermal resistance way
- Press Sensor:Sensitivity 5 uV/V/mmHg, ±2% Impedance 300 to 3000Ω
- Resolution: 1 mmHg
- Unit: mmKg, kPa
- Transducer sites:
Arterial Pressure (ART)
Pulmonary Arterial (PA)
Left Arterial (LAP)
Right Arterial (RAP)
Central Venous Pressure (CVP)
Intracranial Pressure (ICP)
P1/ P2
- Measuring and alarm range:
ART 0 ~ +350mmHg
PA -6 ~ +120 mmHg
CVP/ RAP/ LAP/ ICP -10 ~ +40 mmHg
P1/ P2 -10 ~ +300mmHg
- Accuracy: Static: ±1mmHg or ±2%, whichever is greater (exclusive of transducer)
±4mmHg or ±4%, whichever is greater (inclusive of transducer)
Dynamic: ±4mmHg or ±4%, whichever is greater
Model Configuration
Standard Config.
|
12.1 inch LED, 3/5 Lead ECG, NIBP, SpO2, Pulse Rate,
Temperature, Respiration |
Optional Config. |
1/2 IBP, 2-Temperature, |
Optional Accessories: |
Touch Screen, Printer, Central Monitoring |
Why choose us:
we got production and business certification from Chengdu food and Drug Administration
we got business license from Chengdu City Administration of industry and commerce
we got products certificate of registration and have checked in the website of SFDA
we have accepted the onsite check from SFDA every quarter
we finished the training of laws and regulations in medical equipments field
We have well-established system about material purchasing, employee training,store management,after-sales service process.etc
We adheres the beliefs of "Support Healthy Life", and our company spirits are quality, service,trust,thanksgiving;